← Back to recalls

Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103

Class II
Recalling firm
Accriva Diagnostics, Inc.
Reason for recall
Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.
Affected products (lot / model / serial)
UDI-DI: 10711234103012, Lot(Expiration): B6JCA005(02/28/2027), C6JCA008(03/31/2027), D6JCA011(04/30/2027), K5JCA041(10/31/2026), C6JCA007(03/31/2027)
Units affected
28710
Distribution pattern
Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.
Firm location
San Diego, CA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0317-2027
Recall initiated
September 14, 2026
FDA classification date
October 30, 2026
FDA report date
November 11, 2026
FDA event ID
99863
Last updated on HTM Vibe
November 17, 2026
View on openFDA