← Back to recalls

Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.

Class II
Recalling firm
Cochlear Limited
Reason for recall
Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
Affected products (lot / model / serial)
Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286
Units affected
4 device in the U.S.
Distribution pattern
U.S.: NJ, NY, and PA. O.U.S.: Not provided
Firm location
Macquarie University, Australia
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0349-2027
Recall initiated
September 30, 2026
FDA classification date
November 3, 2026
FDA report date
November 11, 2026
FDA event ID
100098
Last updated on HTM Vibe
November 17, 2026
View on openFDA