← Back to recalls

CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System

Class II
Recalling firm
Wright Medical Technology, Inc.
Reason for recall
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
Affected products (lot / model / serial)
UDI-DI: 00840420109705, Lot: 1826666, Distributed: 18-Aug-2025 through 09-Apr-2026
Units affected
34
Distribution pattern
US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
Firm location
arlington, TN, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-3334-2026
Recall initiated
August 18, 2026
FDA classification date
September 30, 2026
FDA report date
October 7, 2026
FDA event ID
99651
Last updated on HTM Vibe
October 13, 2026
View on openFDA