← Back to recalls

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Class I
Recalling firm
Abiomed, Inc.
Reason for recall
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Affected products (lot / model / serial)
All products below with software version 8.2.2 through 8.5 are affected. 1. Product Code: 0042-0000-CA; GTIN: 00813502011272; Serial Numbers: IC2035, IC9416. 2. Product Code: 0042-0000-EU; GTIN: 00813502011289; Serial Numbers: IC1677, IC3049, IC3099, IC3102, IC4358, IC4329, IC5774, IC5757, IC6580, IC6581. 3. Product Code: 0042-0000-UK; GTIN: 00813502011296; Serial Numbers: IC2023, IC1910, IC2704, IC3325, IC3312. 4. Product Code: 0042-0000-US; GTIN: 00813502010022; Serial Numbers: IC1140. 5. Product Code: 0042-0010-EU; GTIN: 00813502010961; Serial Numbers: IC5403, IC5404, IC5635, IC5636, IC5666, IC5667, IC5801, IC5904, IC5918, IC5921, IC6421, IC6422, IC8683. 6. Product Code: 0042-0010-UK; GTIN: 00813502010978; Serial Numbers: IC8101, IC8102, IC8103, IC8104, IC8544, IC9446, IC9449, IC11440. 7. Product Code: 0042-0040-CA; GTIN: 00813502012958; Serial Numbers: IC10306, IC10312, IC12604, IC12891, IC12895, IC12892, IC12605, IC12894. 8. Product Code: 0042-0040-EU; GTIN: 00813502012217; Serial Numbers: IC9014, IC9022, IC9429, IC10408, IC10425. 9. Product Code: 0042-0040-AU; GTIN: 00813502012972; Serial Numbers: IC10163, IC10387, IC11546, IC11566, IC11490, IC11544, IC12008, IC12054, IC11472, IC12151.
Units affected
62 units
Distribution pattern
International distribution in the country of Australia, Canada, France, Germany, Kuwait.
Firm location
Danvers, MA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
E-Mail
Product type
Devices
Status
Ongoing
Recall number
Z-1591-2026
Recall initiated
February 16, 2026
FDA classification date
March 27, 2026
FDA report date
April 8, 2026
FDA event ID
98393
Last updated on HTM Vibe
July 24, 2026
View on openFDA