Techno-Path Multichem AE Unassayed Level 2 REF: AE622X

Class II
Recalling firm
TECHNO-PATH MANUFACTURING LTD.
Reason for recall
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Affected products (lot / model / serial)
UDI:(01)05391523442817/Lot # 65705262
Units affected
123 units
Distribution pattern
US Nationwide distribution in the state of Massachusetts.
Firm location
Ballina Tipperary, Ireland
Recall type
Voluntary: Firm initiated
How the firm notified customers
E-Mail
Product type
Devices
Status
Ongoing
Recall number
Z-0206-2027
Recall initiated
August 14, 2026
FDA classification date
October 22, 2026
FDA report date
October 28, 2026
FDA event ID
99724