- Recalling firm
- Trividia Health, Inc.
- Reason for recall
- The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
- Affected products (lot / model / serial)
- Device Item/UDI-DI: RF4001-01BK/311917178691, RF4007-01/50428560402, RF4019-01/708820721098, RF4023-01/21292009335, RF4031-01/41220136131, RF4048-01/75450843576, RF4066-01/11822002783, RF4209-01/36800234567, RF4H01-01BK/21292006112, RF4H01-40/21292009793. International: RF4i29-11BK/N/A, RF4i81-11BK/N/A, RF4i82-11BK/21292012229, RF4i82-12BK/21292009984. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX GO Owner's Booklet/ RF4TVH03 Rev 57 and prior, TRUE METRIX GO Owner's Booklet (English)/RF4TVH35 Rev 51 and prior, TRUE METRIX GO Owner's Booklet (Spanish)/RF4TVH35S Rev 51 and prior, TRUE METRIX GO Blood Glucose Monitoring System IFU Trividia House Brand/ RF4ITV16 Rev 55 and prior, TRUE METRIX GO System Instructions for Use (IFU) for Trividia Health UK Limited/ RF4UKT16 Rev 56 and prior, Trividia Health LATAM- TRUE METRIX GO System IFU/ RF4LAT16 Rev 51 and prior, TRUE METRIX GO System Instructions for Use (IFU) for Trividia Health Australia/ RF4AUT16 Rev 57 and prior. Device serial numbers: Prefix "F" followed by an eight-digit numeric sequence.
- Units affected
- 1,407,465
- Distribution pattern
- Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.
- Firm location
- Fort Lauderdale, FL, United States
- Recall type
- Voluntary: Firm initiated
- How the firm notified customers
- Letter
- Product type
- Devices
- Status
- Ongoing
- Recall number
- Z-1548-2026
- Recall initiated
- February 6, 2026
- FDA classification date
- March 12, 2026
- FDA report date
- March 18, 2026
- FDA event ID
- 98317