← Back to recalls

Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd

Class I
Recalling firm
Navajo Manufacturing Company
Reason for recall
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Affected products (lot / model / serial)
UPC: 0-24291-25607-6, UDI-DI: 00024291256076
Units affected
28,457
Distribution pattern
US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.
Firm location
Denver, CO, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-1588-2026
Recall initiated
February 18, 2026
FDA classification date
March 20, 2026
FDA report date
April 1, 2026
FDA event ID
98453
Last updated on HTM Vibe
July 24, 2026
View on openFDA