← Back to recalls

ABIOMED Impella CP Introducer Kit 14 Fr, 13 and 25 cm, Pump Set Code 1000550, containing Kit, 14Fr Introducer, 13cm&25cm, Sterile, Introducer Product Code 0052-3046. heart pump and accessories

Class I
Recalling firm
Abiomed, Inc.
Reason for recall
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Affected products (lot / model / serial)
Pump Set Code 1000550; Introducer Product Code 0052-3046. GTIN: 00813502013474. All current batches in field.
Units affected
846 units
Distribution pattern
Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).
Firm location
Danvers, MA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-2683-2026
Recall initiated
June 11, 2026
FDA classification date
July 10, 2026
FDA report date
July 22, 2026
FDA event ID
98893
Last updated on HTM Vibe
July 28, 2026
View on openFDA