Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.

Class II
Recalling firm
Baxter Healthcare Corporation
Reason for recall
Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
Affected products (lot / model / serial)
Lot Code: Lot numbers: 804084, 804088, and 804089 UDI Number: 00085412475806
Units affected
105675
Distribution pattern
Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia. The countries of Taiwan and Uruguay.
Firm location
Deerfield, IL, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-2262-2026
Recall initiated
April 22, 2026
FDA classification date
May 29, 2026
FDA report date
June 10, 2026
FDA event ID
98852