RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

Class II
Recalling firm
Reflexion Medical, Inc.
Reason for recall
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
Affected products (lot / model / serial)
Software Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.
Units affected
12
Distribution pattern
US: CA, TX, PA, CT, NJ, OR, LA, OH
Firm location
Hayward, CA, United States
Recall type
Voluntary: Firm initiated
Product type
Devices
Status
Ongoing
Recall number
Z-1656-2026
Recall initiated
February 9, 2026
FDA classification date
March 27, 2026
FDA report date
April 8, 2026
FDA event ID
98518