V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.

Class I
Recalling firm
Ventec Life Systems, Inc.
Reason for recall
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Affected products (lot / model / serial)
All Serial Numbers: PRT-01185-000 -UDI: 00855573007877; PRT-01185-001 - UDI: 00850018761161; PRT-01185-002 - UDI: 00850018761154
Units affected
39,568 units
Distribution pattern
Distribution US nationwide including Puerto Rico.
Firm location
Bothell, WA, United States
Recall type
Voluntary: Firm initiated
How the firm notified customers
Letter
Product type
Devices
Status
Ongoing
Recall number
Z-0129-2027
Recall initiated
September 17, 2026
FDA classification date
October 20, 2026
FDA report date
October 28, 2026
FDA event ID
99835